Key Advantages:
• Fiber optics path
• Ambient light rejection technology
• Internal temperature compensation for optics
• Easy to calibrate
• Low calibration solution volume, 50mL minimum
• Provide SS and transparency output(calculated from turbidity)
• RS485 Modbus RTU output
Item |
Technical Specification |
---|---|
Measurement Method |
scattered light at 90 degrees, ISO7027 |
Range | 0.3~1000NTU |
Resolution | 0.1NTU |
Accuracy |
±5% reading or 0.5NTU w.i.g |
Operating temperature |
0~45℃ |
Storage Temperature |
-10~50℃ |
Min. Detection Limit |
0.3NTU |
Warranty |
1 year |
Depth |
IP68, 10m Max |
Power |
DC 5V 5% |
Output |
RS485 and Modbus protocol |
Materials |
Titanium, PVC,PMMA |
Dimensions |
Length 143mm(connector is not included), diameter 16mm |
Flow rate |
< 3 m/s |
Response time |
Minimum 10s T90 |
Field life* |
Sensor 2 years or greater |
Recommended Calibration maintain Frequency* |
Sensor 6 months |
Online Turbidity Sensor (diameter 16mm)
PN: 610212
Broadsensor measures turbidity with an fiber optics sensor. An 860nm
light from the emitter enters the sample via fiber and scatters off
particles in the water. The scattered light at 90 degrees, enters a PD
detector via fiber. This follows the nephelometric technique of
measurement, and values are expressed in nephelometric turbidity
units (NTUs). It follows the ISO7027 standard.
Téléphone:0512-88960831
Retour
Produits recommandés