Key Advantages:

2 wavelength absorption measurement: 254nm and 550nm for turbidity and color compensation.
• SAC measurement used to determine organic loading without reagents.
• Measurement directly in the process, no sample processing or reagent required.
• Fast response, minimum 10 second.
• Cold light source (UV LED) and rugged PD, long life time and low drift.
• Support leakage alarm

Item Technical Specification
Measurement Method
Absorption measurement at 254 nm, DIN 38402 C2, DIN 38404
Range

6mm cuvette gap 0.15~300mg/L equiv. KHP(COD solution)

Resolution 0.1mg/L or 0.01mg/L COD
Repeatability
+/-1% FS equiv. KHP
Operating temperature
0~45℃
Storage Temperature -10~50℃
Clean system
No wiper system
Warranty
1 year
Depth
IP68, 10m Max
Power
5~12V 15mA (normal)
Output RS485 and Modbus protocol
Materials
S316, POM, Nylon, PUR, quartz glass
Dimensions
Length 170.5mm, diameter 34mm
Flow rate
< 3 m/s
Response time
Minimum 10s T90
Field life*
Sensor 3 years or greater
Recommended Calibration Frequency*
Sensor 3 months


Note: *Field life and calibration frequency dependent on site conditions

UV 254 COD Double-Wa

Click to Download

UV 254 Double-Wave COD Sensor

The Broadsensor UV254 sensor is an optical-based absorption and transmittance with UV LED and PD technology sensor. The sensor measures the absorption at 254nm and 550nm at the same time to compensate the SAC measurement for interferences from turbidity and color.